Infection risk in cardiothoracic surgery due to contaminated heater cooler units

Introduction

When several cases of severe bacterial infections with Mycobacterium chimaera in a Swiss hospital became public so called heater cooler units (HCU) used during open heart surgery were suspected to be the origin of the pathogens.

Objectives

After confirming that the tanks of the HCU used in our facility were contaminated we attempted to identify potential transmission ways.

Methods

To enable a timely risk assessment and due to the fact that Mycobacterium chimaera are very slow growing organisms we used non-fermenters as a surrogate parameter for water associated pathogens. In an experimental setting we conducted air sampling in different distances to the operating HCU. In addition air was sampled in the area of the operating table while HCU were placed at their usual location in the operating room. This was conducted either when HCU were operating or switched off. Air (200 L in 2 minutes) was collected by air sampler MBASS30LKS100 and conducted over microbiological plates for culturing.

Results

Non-fermenters could be identified in up to three meters distance to the operating HCU. As long as the HCU were not operating cultures from the area of the operating table remained negative for non-fermenters. Once the HCU were started cultures showed considerable growth of non-fermenters.

Conclusion

Considering our results pathogens from contaminated HCU tanks are disseminated via aerosol formation. Thus there is a potential way of transmission to patients undergoing open heart surgery. So far the extend of the problem is unknown which is why we would like to encourage other hospitals to initiate similar examinations. Our results suggest that HCU should be banned from the operating room. As they are also used in intensive care units this should be further evaluated as well.

Disclosure of interest

None declared.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Götting, T., Jonas, D., Scheibert, D. et al. Infection risk in cardiothoracic surgery due to contaminated heater cooler units. Antimicrob Resist Infect Control 4 (Suppl 1), P73 (2015). https://doi.org/10.1186/2047-2994-4-S1-P73

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2047-2994-4-S1-P73

Keywords