Fig. 2 | Antimicrobial Resistance & Infection Control

Fig. 2

From: The surgical wound in infrared: thermographic profiles and early stage test-accuracy to predict surgical site infection in obese women during the first 30 days after caesarean section

Fig. 2

Bench-top black body instrument (a) set to 38 °C showing equivalent temperature measured with the camera lens positioned in front of the heat source and with pixel value obtained (at cross-hair) of 38 °C (b). Graph (c) shows results for reliability of the thermal camera across a temperature range (31-46 °C) revealing consistency between camera and black body source to within + 0.1o to 0.2 °C. Distance of camera lens from black body source has minimal effect on temperature readings (d)